Maintenance
82
|
Dionex IC Pure system
Thermo Scientific
Chapter 13
Disinfection
Required for disinfection is the disinfection adapter (item no: 50133431) purchase separately.
Use the following disinfecting agent:
Cleaning Solution, 1 syringe, item no.: CMX 25.
The Dionex IC Pure system can only be disinfected as a 'stand-alone' process and cannot be
performed while in operation with an IC system.
Perform the following steps if an ion chromatograph system is connected to the Dionex IC Pure
system:
Switch off the IC system and remove the level switch cable for the Dionex Integrion and Dionex
ICS-2100, Dionex ICS-4000, Dionex ICS-5000
+
, Dionex ICS-5000 or Dionex ICS-3000 system
from the Dionex IC Pure system.
Disconnect the tubes for the IC system from the connections at the Dionex IC Pure system.
Note
To ensure effective disinfection of your unit fill the internal 5 L or external 20 L tank 80%
with water. You can read off the level of both tanks using the additional tank indicator on the unit.
Always wear protective gloves when handling the cleaning solution.
Always wear safety goggles when handling the cleaning solution.
CAUTION
If your skin should come into contact with a chlorine product, immediately rinse it
with ample, fresh water. If your eyes come into contact with the disinfecting agent, immediately rinse
them with ample, fresh water and immediately contact a physician.
Summary of Contents for dionex ic pure water
Page 10: ......
Page 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Page 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Page 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Page 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Page 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Page 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Page 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Page 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Page 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Page 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Page 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Page 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Page 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Page 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Page 127: ...Thermo Scientific IC Pure 117 ...