Dionex Easion Operator’s Manual
48
Doc. 155028-01 08/20
The table below lists the Dionex Easion-related error messages, as well as their
default severity levels. For troubleshooting assistance, go to the page indicated in
the table.
If this error occurs when you turn on the Dionex Easion power, you can
disregard it. If the error occurs at other times, it should be resolved.
To troubleshoot:
Refer to the suppressor manual for troubleshooting procedures.
This error occurs if you try to calibrate the flow rate while the pump is off, or
while the pump is on but the flow rate is not 1 mL/min.
To troubleshoot:
Be sure to click
Start Calibration
on the Wellness panel with the flow rate
calibration instructions. This automatically starts the pump and sets the flow
rate to 1 mL/min. For flow rate calibration instructions, see
.
Audit Trail Error Message
Default
Severity Level
See
Warning
Warning
Error
Module data buffer overflow. Data may have been
lost.
Warning
Warning
Warning
Pump pressure slope calibration error.
Warning
Remaining eluent below 100 mL.
Error
Remaining eluent below 200 mL.
Error
The system pressure has exceeded the high
pressure limit.
Abort
The system pressure is below the low pressure limit.
Abort
Conductivity exceeds limit.
Flow rate calibration error.