2 • Description
Doc. 155028-01 08/20
23
The conductivity cell has two detection limit ranges: 0 to 500
S (the
default range) and 0 to 10,000
S. The range to select depends on the
expected detector readings for the application. The default range is
appropriate for most applications.
Select the detection limit range in the Chromeleon Instrument
Configuration Manager (in the Dionex Easion Properties dialog box). For
more information, refer to the Chromeleon Help.
2.4
Chromeleon Software
The Chromeleon 7 Chromatography Data System (version 7.2.10 MUa and later)
is used to control the Dionex Easion. Chromeleon provides complete instrument
control, data acquisition, and data management.
2.4.1
The Chromeleon ePanel Set
The Chromeleon ePanel Set provides access to detailed status and control
functions for each system component (pump, detector, and so on). The
Home tab of the ePanel Set includes system status information, a signal
plot, and controls for the most commonly used system functions. For
instructions on how to connect to the ePanel Set, see
.
Figure 2-9. Conductivity Cell and Dionex DS5 Detection Stabilizer
From suppressor
ELUENT OUT port
To regenerant reservoir
Connector to component
mounting panel
DS
5 D
ete
cti
on
St
ab
iliz
er
Conductivity Cell
Cell Inlet
Cell Outlet