Dionex Easion Operator’s Manual
82
Doc. 155028-01 08/20
5. Disconnect the orange
CELL OUT
line from the union that connects it to the
black
CELL OUT
line.
6. Use an Allen wrench to remove the two screws on the top of the cell housing.
7. Remove the cell by unplugging it from the component mounting panel.
8. Plug the new conductivity cell (P/N 067761) into its mounting location.
9. Replace the screws on the top of the cell housing and tighten.
10. Attach the orange
CELL OUT
line from the cell to the union attached to the
black
CELL OUT
line.
11. Attach the
ELUENT OUT
line to the
ELUENT OUT
port on the suppressor.
12. Close the Dionex Easion front door.
13. Reconnect the power cord and turn on the Dionex Easion power.
Figure 5-10. Conductivity Cell Assembly
to suppressor
Connector to component
DS
5 D
ete
cti
on
St
ab
iliz
er
ELUENT OUT
mounting panel
CELL OUT (orange)
to CELL OUT (black)