Dionex Easion Operator’s Manual
44
Doc. 155028-01 08/20
4. Verify that the injection valve is in the Load position. If it is not, click
Load
on the Chromeleon ePanel (see
5. Insert a syringe into the sample loading port on the Dionex Easion
front door and pull out the plunger to draw the sample into the
injection valve. Draw at least five times the sample loop volume.
NOTE The line from the sample to port
W (6)
on the
injection valve must be filled before the sample
loop can be filled.
6. Use one of the following methods to inject the sample onto the
column.
•
Manually: Click
Inject
on the Chromeleon ePanel. For more
information about manual sample processing, see
.
•
Automatically: Include the Inject command in a Chromeleon
instrument method (see
). For more information
about automatic sample processing, see
3.11.6 Example Chromeleon Commands for Loading and
Injecting Samples
The following example shows commands for loading and injecting
samples using a Dionex AS-DV.
Example Dionex AS-DV Instrument Method Commands
Sampler.LoadPosition
;Switch the valve to Load.
DeliverSample
;Deliver sample to the valve.
0.000
Autozero
;Zero the baseline.
Wait
CycleTimeState
;Wait for cycle time (if any).
Inject
;Switch the valve to Inject.
ECD_1.AcqOn
;Start data acquisition.
14.000
ECD_1.AcqOff
;Stop data acquisition.