4 • Troubleshooting
Doc. 155028-01 08/20
49
This error occurs if the injection valve fails to actuate within 1 second of
being toggled.
To troubleshoot:
1. If a sequence is being executed, click
Stop
on the Chromeleon ePanel to
cancel the current injection and stop the sequence.
2. Turn off the Dionex Easion power briefly and then restart.
3. Try to toggle the valve from Load to Inject by clicking
Inject
on the
Chromeleon ePanel.
4. If the problem persists, contact Technical Support for Dionex products for
assistance.
This error can be caused by various electronics-related problems.
To troubleshoot:
Contact Technical Support for Dionex products for assistance. The Dionex
Easion electronics components cannot be serviced by the user.
This error indicates a problem in the pump controller electronics.
To troubleshoot:
Contact Technical Support for Dionex products for assistance. The Dionex
Easion electronics components cannot be serviced by the user.
This error indicates a problem in the pump controller electronics.
To troubleshoot:
Contact Technical Support for Dionex products for assistance. The Dionex
Easion electronics components cannot be serviced by the user.
Load/inject valve error.
Module data buffer overflow. Data may have been lost.
Pump motor lost control.
Pump pressure hardware error.