7
•
Maintenance and Service
Page 182
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
2. Pull the nebulizer straight out of the nebulizer opening.
NOTICE
Avoid the nebulizer tip from touching the sides of the
nebulizer opening, as this can damage the tip.
Figure 33: Removing the nebulizer
3. Install the dust seal plug to the opening for the nebulizer.
4.
If the nebulizer is to be transported
Pack the nebulizer:
a) Carefully install the protective tubing to the nebulizer spray
pipe. Do not to touch the nebulizer tip with the protective
tubing, as it can damage the tip.
b) Install the plug to the nebulizer flow inlet.
Figure 34: Nebulizer with protective tubing and plug on the flow inlet
c) Pack the nebulizer in its original packaging that it was shipped in.
To re-install the nebulizer, follow the instructions in the
Installation
Guide
that is shipped with the nebulizer.
Summary of Contents for Corona Veo
Page 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Page 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...