7
•
Maintenance and Service
Page 192
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
7.9
Stream-Switching Valve (Corona Veo RS only)
This section applies only for Corona Veo RS detectors.
7.9.1
Disassembling and Inspecting the Valve
Tools required
Hexagon socket-head wrench, size 9/64"
Preparations
1. Flush the detector with the detector gas flow being turned on. For
recommendations on flushing the detector, see
section 7.4.3
Flushing the Detector between Analyses, page 172
2. Turn off the pump flow and then the gas flow:
a) Turn off the pump flow to the detector. Wait at least 5 minutes.
During the waiting period, the detector drains any remaining
mobile phase to waste.
b) Turn off the gas flow on the detector.
c) Wait until the system pressure is down to zero before you
continue.
d) Turn off the gas supply to the detector.
3. Turn off the detector with the main power switch.
4. Disconnect the capillaries from the stream-switching valve.
Summary of Contents for Corona Veo
Page 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Page 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...