5
•
Installation
Page 80
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
3. Check the length of the gas tubes.
The gas tubes should go straight to the detector. If the gas tubes are
longer than required, use a tubing cutter to cut the tubes to the
required length.
♦
Make sure that the cut is at a right angle to the length of the gas
tubes.
♦
A proper cut is especially important to provide a good
connection to the push-in fitting on the gas inlet. Slanted or
angular cuts can lead to gas leaks.
TIP
If you wish to use a different gas inlet tubing with metric
dimensions, use the inch-to-metric adapter to connect the tubing to
the gas inlet of the detector.
4. Connect the gas inlet tubing to the push-in fitting of the gas inlet on
the rear panel of the detector.
5. Connect the gas exhaust tubing to the push-in fitting of the exhaust
output of the detector.
Figure 9: Gas inlet and exhaust on the detector
No.
Description
1
Gas exhaust tubing
2
Gas inlet tubing
Summary of Contents for Corona Veo
Page 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Page 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...