3
•
Detector Overview
Page 44
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
3.4
Nebulizer
The FocusJet nebulizer is the liquid flow inlet of the charged aerosol
detector. Eluate from the column flows into the nebulizer and merges
with filtered gas that is supplied to the nebulizer. Due to the concentric
flow design of the nebulizer, an aerosol is produced within the spray
chamber.
Figure 3: FocusJet nebulizer
3.5
Gas Filter Assembly
An assembly of two gas filters on the inner right side panel of the
detector filters the supplied gas internally before it is used in the
detector. The gas filters consist of a charcoal filter and a high-efficiency
particulate air (HEPA) filter. Gas supplied through the gas inlet of the
detector is first filtered in the charcoal gas filter (bottom filter) and then
in the HEPA gas filter (top filter).
Sufficient filtering of the supplied gas, both externally and internally,
ensures that the gas is clean and free from micro particles.
Summary of Contents for Corona Veo
Page 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Page 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...