6
•
Operation
Page 112
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
6.4.2.1
Diagnostics Screen
The
Diagnostics
screen displays real-time information and values with
regard to factory-set and internally calibrated parameters.
If no errors or warnings are active, the
Status
is
Normal
.
Figure 20: Diagnostics screen (here: Corona Veo RS)
Diagnostic
parameter
Description
Typical Values
Flow Ratio
Indicates the ratio of nebulizer gas pressure to
the charging gas pressure (low, normal or high),
calibrated at factory
Normal
Ion Trap
Voltage that is used to remove highly mobile gas
ions
18 V - 20.9 V
Gas Pressure
Indicates the inlet pressure from the gas supply
40 - 65.0 psig
± 0.1 psig (276 -
448 kPa, 2.8 -
4.5 bar)
Evaporation
Temp
•
Indicates if the evaporation tube heater is
operating
•
Displays the actual evaporation temperature
On, 35 °C
Charger
Voltage
Indicates the applied voltage to the corona
needle to charge the (secondary) ion gas stream
1.8 kV - 3.0 kV
Charger
Current
Indicates the current required to produce ions in
the (secondary) gas stream
0.99 µA - 1.01 µA
Summary of Contents for Corona Veo
Page 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Page 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...