3.2
Operating Principle
The detector is designed for charged aerosol detection with full
compatibility in HPLC and UHPLC applications. In charged aerosol
detection, an aerosol of dried analyte particles is formed after
nebulization of column eluate. Electrical charge is imparted to these
dried particles. The detector then measures the charge. Measured
charge is proportional to the amount of analyte in the sample.
The following picture shows the interior components of the detector,
and illustrates how the detector operates:
Figure 1: Operating principle of the detector
No. Description
No. Description
1
Mobile phase from column
8
Mixing chamber
2
Nebulizer
9
Charger gas inlet
3
Gas inlet
10
Corona charger
4
Gas filters
11
Ion trap
5
Spray chamber
12
Electrometer
6
Drain pump
13
Electrometer board
7
Evaporation tube
14
Gas exhaust
Mobile phase from the column enters the detector (no. 1) and flows to
the nebulizer (no. 2). Pressurized gas streams through the gas inlet
(no. 3) and passes through the charcoal gas filter and then the HEPA gas
filter (no. 4). In the nebulizer (no. 2), the filtered gas and the mobile
Summary of Contents for Corona Veo
Page 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Page 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...