Calibrating the PDA Detector from the ChromQuest Data System
Thermo Scientific
Accela PDA Detector Hardware Manual
75
Performing a Wavelength Calibration from the ChromQuest Data System
The alignment of the spectrum on the diode array depends upon the physical alignment of
various components of the optical bench. The alignment can become offset if the detector is
sharply jolted, in shipping, for example. Such bumps and jars can slightly change the
wavelength of light reaching the photodiode array. The automated wavelength calibration
determines the wavelength accuracy of the detector and uses the detector’s wavelength
algorithm to correct any misalignment.
Y
To perform a wavelength calibration with ChromQuest
1. Open the Accela PDA Diagnostics dialog box (see
“Accessing the Direct Controls for the
2. Click the
Calibration
tab.
The Calibration page appears (see
Figure 62.
Calibration page of the Accela Diagnostics dialog box