ChromQuest Diagnostics for the PDA Detector
Adjusting the Light Throughput with the ChromQuest Data System
72
Accela PDA Detector Hardware Manual
Thermo Scientific
shows a saturated array in the UV region.
Figure 59.
Saturated diode array (UV region)
shows a diode array that is saturated in both the UV and visible regions. The
configuration is set to a 20 Hz diode array scan rate and the attenuators are completely open.
Figure 60.
Saturated diode array (UV region and visible region)
4. After you finish adjusting the attenuators, close the front doors of the detector.
The diode array is saturated in the UV region.
The diode array is saturated in the UV region and the visible region.