Configuration
Thermo Foundation Instrument Configuration
28
Accela PDA Detector Hardware Manual
Thermo Scientific
Thermo Foundation Instrument Configuration
To control the PDA detector from the Xcalibur data system, add it to the configured devices
list in Thermo Foundation, and then specify its stack number and diode array scan rate.
Y
To configure your instrument in Thermo Foundation
1. Open the Instrument Configuration application.
For Xcalibur 2.1 or higher, from the computer desktop, choose
Start > All Programs >
Thermo Foundation 1.0 > Instrument Configuration
.
The Thermo Foundation Instrument Configuration window appears (see
Figure 22.
Thermo Foundation Instrument Configuration window
2. In the Available Devices list, double-click each of your Accela devices.
As you double-click a device in the Available Devices list, it appears in the list of
configured devices.