4
Xcalibur Diagnostics for the PDA Detector
Calibrating the PDA Detector from the Xcalibur Data System
50
Accela PDA Detector Hardware Manual
Thermo Scientific
b. In the Wavelength Calibration area, click
Calibrate
.
The preconditions page of the wavelength calibration wizard appears (see
Figure 41.
Preconditions page
c.
Read the preconditions, and determine if they have been met:
• If the preconditions have been met, click
Next
to proceed with the calibration.
• If the preconditions have not been met, click
Cancel
to close the wizard and
prepare the PDA detector for calibration.
4. On page 2 of the wavelength calibration wizard (see
), do the following:
a. Select a peak set from the list.
The peak set should span the wavelengths you use under normal operating
conditions.
shows the selection of the Holmium Oxide 5 peak set. The
detector uses this wavelength list to perform the wavelength calibration.
Note
You can click Cancel on any page of the Wavelength Calibration
wizard to stop the calibration process.
Note
Xcalibur has four calibration files to choose from. For example, the
Holmium Oxide UV file contains five wavelengths in the UV region while the
other files use sets of wavelengths from both the UV and Visible wavelength
regions. The holmium oxide absorbance maxima are selected from a spectrum
published in “Holmium Oxide Solution Wavelength Standard from 240 to
640 nm - SRM 2034 (NIST Special Publication 260-54).”