6
Routine Maintenance
Replacing the Lamps
Thermo Scientific
Accela PDA Detector Hardware Manual
93
12. To remove the deuterium lamp assembly:
a. To disconnect the lamp cable, squeeze the connector latch, and then gently pull the
connector free from the receptacle on the lamp compartment wall (see
b. Loosen but do not remove the two Phillips head screws by turning them
approximately four turns.
c.
Lift and gently twist the mounting flange counterclockwise to free the lamp
(see
Figure 76.
Lamp connections
Deuterium lamp
connector
Tungsten lamp
connector