System Description
Peripherals Support System
8-8
48iQTL Instruction Manual
Thermo Scientific
The peripherals support system operates those additional devices that are
needed, but do not require special feedback control or processing. These
components are connected to a Peripherals Support Board (PSB).
The PSB microprocessor regulates the timing and health of such devices,
and generates registers that interact with the higher level system controls.
The chassis fan provides air cooling of the active electronic components.
The Step POL board provides high/low outputs for continuous operation
or on/off states. The Step POL board contains the basic circuitry to provide
a programmable load to passive devices, either continuously, or on user or
automated command. In the iQ Series instruments, the pump, solenoids,
etc., are controlled off of the Step POL board from commands generated
via the PSB.
Internal vacuum pump for generating air/sample through the instrument.
Optional solenoid valves for switching between sample, zero, and span
gases, and other optional components.
The sample/zero valve, located downstream of the sample bulkhead and the
permeation dryer, is a 3-way solenoid that routes sample flow either
through or around the CO scrubber. During a zero adjustment, zero check
or the scrubber test, the valve activates to route gas through the scrubber.
The Flow/Pressure DMC is used measure instrument pressures that assure
proper flow regulation and for sample pressure within the measurement
bench for pressure corrections and compensation.
The DMC includes two pressure sensors that read 0
860 mmHg. These
sensors are used with the coupled restricting capillary for flow control along
with the downstream sample pump. The pressure differential determines
the flow through the capillary. The upstream pressure is the measurement
bench pressure, while the downstream pressure is at the pump inlet
pressure.
Peripherals
Support System
Fan
Step POL Board
Sample Pump
Solenoid Valve
Panel (optional)
Sample/Zero Valve
Flow/Pressure
DMC
Summary of Contents for 48iQTL
Page 1: ...48iQTL Instruction Manual Trace Level Carbon Monoxide Analyzer 119014 00 27Sep2018 ...
Page 2: ......
Page 32: ...Operation Instrument Display 3 8 48iQTL Instruction Manual Thermo Scientific ...
Page 142: ......
Page 153: ...Calibration Calibration Thermo Scientific 48iQTL Instruction Manual 4 11 ...
Page 230: ......
Page 240: ......
Page 266: ......
Page 278: ......
Page 279: ......