Servicing
DMC Optical Bench
Thermo Scientific
48iQTL Instruction Manual
7-43
Use the following procedure to remove the pre-amplifier/detector assembly.
Equipment required:
Phillips drive, #1
Hex drive, 7/64
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Unplug preamp and source IR cable assembly.
Figure 7–46.
Preamp Detector Cover Removal pt 1
3.
Using a #1 Phillips drive, unfasten two #4-40 pan head screws.
4.
Lift preamp cover up.
Pre-Amp/Detector
Assembly Removal