Troubleshooting
Troubleshooting Guide
Thermo Scientific
48iQTL Instruction Manual
6-3
Problem
Possible
Cause
Action
Dirty fan filter
Clean or replace filter.
Overheating PCB Locate defective PCB reporting the error
and replace if needed.
Alarm – Bench Temperature
Defective bench
heater
Check 10K ohm thermistor and replace
heater if bad.
Alarm – Pressure
High bench
pressure
indication
Check the pump for a tear in the
diaphragm; replace with pump repair kit
if necessary. Refer to “
Maintenance
”
chapter in this manual.
Check that capillaries are properly
installed and o-rings are in good shape.
Replace if necessary.
Check flow system for leaks.
Alarm – Flow
Flow low
Check sample capillary for blockage.
Replace as necessary.
If using sample particulate filter, make
sure it is not blocked. Disconnect sample
particulate filter from the sample
bulkhead. If flow increases, replace the
filter.
Perform a leak test as decribed on page
5-5
.
Flow high
When delivering zero air or gas to the
instrument, use an atmospheric dump.
Flow = 0 LPM
Check that Step POL board #1 has both
dip switch settings of SW2 off (both
facing the rear of the instrument). Verify
the pump is plugged into the Step POL
board.
Worn
Diaphragm
Rebuild pump every 12 months or as
needed.
Alarm – Board
Communication
Cable
connection
Check that DMC cable is connected
properly. Reseat if needed.
Defective
DMC
PCB
Replace DMC board.
Alarm – Power Supply
Cable
connection
Check that DMC cable is connected
properly. Reseat if needed.
Defective
component
Check for other alarms, as it is possible
that another component of that DMC is
drawing too much current.
Defective
DMC
PCB
Replace DMC board.
Summary of Contents for 48iQTL
Page 1: ...48iQTL Instruction Manual Trace Level Carbon Monoxide Analyzer 119014 00 27Sep2018 ...
Page 2: ......
Page 32: ...Operation Instrument Display 3 8 48iQTL Instruction Manual Thermo Scientific ...
Page 142: ......
Page 153: ...Calibration Calibration Thermo Scientific 48iQTL Instruction Manual 4 11 ...
Page 230: ......
Page 240: ......
Page 266: ......
Page 278: ......
Page 279: ......