Operation
Settings
Thermo Scientific
48iQTL Instruction Manual
3-83
The Communications screen allows the user to set TCP/DHCP
parameters, Serial settings, Analog I/O and Digital I/O, Email Server, and
Instrument ID. Buttons are grayed out if not selected in
Settings>
Configuration
.
Home Screen>Settings>Communications
The Communications screen contains the following information:
●
Wired TCP/DHCP:
Settings for communicating with the instrument
through wired Ethernet.
●
Serial RS-232/485:
Settings for communicating with the instrument
through RS-232/485 protocol. This is only visible if selected in
Settings>Configuration>Communications Board.
●
Analog I/O:
Settings for communicating with the instrument through
analog I/O settings. This is only visible if selected in
Settings>Configuration>Analog I/O.
●
Digital I/O:
Settings for communicating with the instrument through
digital I/O settings. This is only visible if selected in
Settings>Configuration>Digital I/O.
●
Email Server (SMTP):
Settings for communication with email.
●
Instrument ID:
Allows the user to edit the instrument identification
number (ID). The ID is used to identify the instrument when using
protocols to control the instrument or collect data. It may be necessary
to edit the ID number if two or more of instruments of the same model
are connected to one computer. Valid instrument ID numbers are from
0 to 127. The 48iQTL has a default instrument ID of 1.
●
Bayern Hessen Settings:
User selects the registers (measured values) that
the instrument should respond with, in response to a valid data query
(DA) command.
Communications
Summary of Contents for 48iQTL
Page 1: ...48iQTL Instruction Manual Trace Level Carbon Monoxide Analyzer 119014 00 27Sep2018 ...
Page 2: ......
Page 32: ...Operation Instrument Display 3 8 48iQTL Instruction Manual Thermo Scientific ...
Page 142: ......
Page 153: ...Calibration Calibration Thermo Scientific 48iQTL Instruction Manual 4 11 ...
Page 230: ......
Page 240: ......
Page 266: ......
Page 278: ......
Page 279: ......