Operation
Settings
Thermo Scientific
48iQTL Instruction Manual
3-81
●
Units:
Displays the units of measurement.
●
Down:
●
Test Gas Concentration:
Displays the test gas concentration as
measured by the instrument prior to being scrubbed.
●
Current Concentration:
Displays the current measured CO
concentration.
●
Scrubber Efficiency:
Displays the calculated scrubber efficiency at
the end of the test.
●
Start Cal:
When Start is pressed, the scrubber efficiency test is
initiated. Please allow approximately 15 minutes for the test to
complete. During the first part of the test, the test gas is routed
directly to the bench, bypassing the scrubber in order to establish
the test gas concentration as measured by the instrument. During
the second part of the test the test gas is routed through the
scrubber prior to being measured by the instrument. The scrubber
efficiency is calculated at the completion of this part of the test.
The user can stop the test pressing the Stop button.
●
End Cal:
When Stop is pressed, the scrubber efficiency test is
interrupted.
●
Last Test Completed:
Displays the date of the last successfully
completed scrubber efficiency test.