Thermo Scientific
48iQTL Instruction Manual
6-1
Chapter 6
Troubleshooting
This chapter presents guidelines for diagnosing analyzer failures, isolating
faults, and includes recommended actions for restoring proper operation.
Read the safety precautions in Appendix A, “
Safety
” before performing any
actions listed in this chapter.
Table 6–1
provides general troubleshooting information for the common
platform and indicates the checks that you should perform if you
experience an instrument problem. It also lists 48iQTL specific
troubleshooting information and alarm messages you may see on the
graphics display and provides recommendations about how to resolve the
alarm condition.
Table 6–1.
48iQTL Troubleshooting Guide
Problem
Possible
Cause
Action
Instrument does not start
(LEDs on front panel do not
come on and display is
blank)
No power
Verify that the power cord is plugged in,
power is available and that it matches
the voltage and frequency configuration
of the instrument.
Fuse is blown or
missing
Disconnect power and check fuses with a
volt meter.
Bad switch or
wiring
connection to
switch
Check for 24V @ J9 on the Backplane
board (middle pins). Check all wiring
connections.
Front panel display does not
start (LEDs on front panel are
off)
Disconnected
ribbon cable
Power down and evaluate connections of
display ribbon cable.
Front panel display does not
start (LEDs on front panel are
on)
Defective
Display
Connect to the instrument using ePort.
Select “Remote Interface”. If normal GUI
is displayed, replace defective display.
Front panel display stays
white after power up (LEDs
on front panel are on)
Unseated or
missing Micro
SD card
Power off, re-seat Micro SD or install if
missing.
Safety Precautions
Troubleshooting
Guide
Summary of Contents for 48iQTL
Page 1: ...48iQTL Instruction Manual Trace Level Carbon Monoxide Analyzer 119014 00 27Sep2018 ...
Page 2: ......
Page 32: ...Operation Instrument Display 3 8 48iQTL Instruction Manual Thermo Scientific ...
Page 142: ......
Page 153: ...Calibration Calibration Thermo Scientific 48iQTL Instruction Manual 4 11 ...
Page 230: ......
Page 240: ......
Page 266: ......
Page 278: ......
Page 279: ......