Servicing
DMC Optical Bench
7-30
48iQTL Instruction Manual
Thermo Scientific
Use the following procedure to remove the optical bench from the
instrument case.
Equipment required:
Phillips drive, #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug DMC cable J3 (top). Unplug heater cable J7. Unplug case fan
cable J12.
3.
Disconnect plumbing as shown.
4.
Using a #2 Phillips drive, unfasten six captive hardware.
Figure 7–35.
DMC Removal from Instrument