Servicing
Optional Manifold Replacement
Thermo Scientific
48iQTL Instruction Manual
7-55
Use the following procedure to replace the manifold.
Equipment required:
Hex wrench, 9/16
Hex drive, 9/64
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug three electrical connections (J5, J6, and J8) from the Step POL
board 2.
Figure 7–59.
Replacing the Manifold pt 1
3.
Unfasten three nuts. Remove the nuts, front and back ferrules as shown
from span, zero in, sample back panel (
Figure 7–60
).
Optional
Manifold
Replacement