Operation
Calibration
3-24
42iQHL Instruction Manual
Thermo Scientific
The Adjust NO and NO
x
Background screens are used to adjust the
instrument zero background. Before making an adjustment, be sure the
analyzer samples zero air for at least 5 minutes.
It is important to note the averaging time when calibrating. The longer the
averaging time the more precise the calibration results. To achieve
maximum precision, allow the instrument to stabilize each time input gas is
changed and set the averaging time to 300-second averaging.
Home Screen>Calibration> Advanced Calibration>Manual
Calibration>Adjust Backgrounds>Adjust NO Background
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust Backgrounds>Adjust NO
x
Background
The Adjust NO and NO
x
Background screens contain the following
information:
●
Adjust NO Background:
User manually adjusts zero NO background.
●
Adjusted NO Concentration:
Read only. Displays what the NO
concentration value will become when the calibrate button is pressed.
Adjust NO and NO
x
Background