Operation
Settings
Thermo Scientific
42iQHL Instruction Manual
3-107
The Analog I/O screen allows for configuring the analog inputs/outputs.
This is only visible if selected in Settings>Configuration>
Analog I/O
.
Home Screen>Settings>Communications>Analog I/O
The Analog I/O screen contains the following information:
●
Analog In:
Allows the user to view and calibrate voltage inputs from
external devices.
●
Analog Out (Voltage):
Allows the user to view voltage outputs.
●
Analog Out (Current):
Allows the user to view current (mA) outputs.
●
Analog Out Under/Over Range Enabled/Disabled:
Allows the user to
select whether or not the analog outputs are allowed to exceed the
selected output range.
●
Analog Out Calibration:
Allows the user to calibrate the analog out
voltage and current to account for varying loads attached to the
terminals.
Analog I/O