Servicing
DMC Ozonator
7-48
42iQHL Instruction Manual
Thermo Scientific
Figure 7–49.
Replace Transformer
Use the following procedure to replace the ozonator and transformer board.
Equipment Required:
Phillips drive, #2
1.
Remove DMC Ozonator as described in “
DMC Ozonator
” in this
chapter.
2.
Unplug all electrical from HPVS board, ozonator (J1), flow switch (J5)
converter DMC (J2) and flow pressure board (J4).
3.
Using a #2 Phillips drive, unfasten four #6-32 pan head screws.
Ozonator and
Transformer Board
Replacement