Operation
Settings
3-124
42iQHL Instruction Manual
Thermo Scientific
The Configuration screen allows the user to enable optional features. If an
option is disabled, the corresponding buttons will be greyed out and the
screens will not be available.
Use the
and
buttons to select the variables. Next, press the
Commit
Changes
button to save selections. Yellow buttons indicate that the
variable is selected. More than one can be chosen.
Home Screen>Settings>Configuration
The Configuration screen contains the following information:
●
Converter Molybdenum:
Select if molybdenum converter is used.
●
Converter Stainless Steel:
Select if stainless steel converter is used.
●
Zero Span Assembly:
Enables zero/span valves.
●
Oxygen Sensor:
Enables O
2
sensor option.
●
NH
3
Scrubber:
Removes NH
3
from sample.
●
Bypass Flow:
Enables bypass flow option
.
●
Communications Board:
Enables RS-232 or RS485 communications
board option.
●
Predictive Diagnostics:
Enables predictive diagnostics.
●
Analog I/O:
Enables analog I/O option.
●
Digital I/O:
Enables digital I/O option.
Configuration
Summary of Contents for 42iQHL
Page 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Page 2: ......
Page 18: ......
Page 28: ......
Page 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Page 166: ......
Page 192: ......
Page 272: ......
Page 308: ......
Page 320: ......
Page 321: ......