Operation
Settings
3-102
42iQHL Instruction Manual
Thermo Scientific
The Communications screen allows the user to set TCP/DHCP
parameters, Serial settings, Analog I/O and Digital I/O, Email Server, and
Instrument ID. Buttons are grayed out if not selected in
Settings>
Configuration
.
Home Screen>Settings>Communications
The Communications screen contains the following information:
●
Wired TCP/DHCP:
Settings for communicating with the instrument
through wired Ethernet.
●
Serial RS-232/485:
Settings for communicating with the instrument
through RS-232/485 protocol. This is only visible if selected in
Settings>Configuration>Communications Board.
●
Analog I/O:
Settings for communicating with the instrument through
analog I/O settings. This is only visible if selected in
Settings>Configuration>Analog I/O.
●
Digital I/O:
Settings for communicating with the instrument through
digital I/O settings. This is only visible if selected in
Settings>Configuration>Digital I/O.
●
Email Server (SMTP):
Settings for communication with email.
●
Instrument ID:
Allows the user to edit the instrument identification
number (ID). The ID is used to identify the instrument when using
protocols to control the instrument or collect data. It may be necessary
to edit the ID number if two or more of instruments of the same model
are connected to one computer. Valid instrument ID numbers are from
0 to 127. The 42iQHL has a default instrument ID of 1.
●
Bayern Hessen Settings:
User selects the registers (measured values) that
the instrument should respond with, in response to a valid data query
(DA) command.
Communications
Summary of Contents for 42iQHL
Page 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Page 2: ......
Page 18: ......
Page 28: ......
Page 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Page 166: ......
Page 192: ......
Page 272: ......
Page 308: ......
Page 320: ......
Page 321: ......