Operation
Settings
Thermo Scientific
42iQHL Instruction Manual
3-89
●
High Range NO Setting:
Sets the high range NO concentration range
for the analog output when in dual or auto range mode.
●
Low Range NO Setting:
Sets the low range NO concentration range for
the analog output when in dual or auto range mode.
●
NO
2
Range Setting:
Sets the NO
2
concentration range for the analog
outputs when in single range mode.
●
High Range NO
2
Setting:
Sets the high range NO
2
concentration range
for the analog output when in dual or auto range mode.
●
Low Range NO
2
Setting:
Sets the low range NO
2
concentration range
for the analog output when in dual or auto range mode.
●
NO
x
Range Setting:
Sets the NO
x
concentration range for the analog
outputs when in single range mode.
●
High Range NO
x
Setting:
Sets the high range NO
x
concentration range
for the analog output when in dual or auto range mode.
●
Low Range NO
x
Setting:
Sets the low range NO
x
concentration range
for the analog output when in dual or auto range mode.
Settable ranges according to unit selection include:
ppb 10000
5000000 ppb
ppm 10
5000 ppm
% 0.001
0.5 %
μg/m
3
20000
7500000 μg/m
3
mg/m
3
20
7500 mg/m
3
g/m
3
0.02
7.5 g/m
3
Summary of Contents for 42iQHL
Page 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Page 2: ......
Page 18: ......
Page 28: ......
Page 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Page 166: ......
Page 192: ......
Page 272: ......
Page 308: ......
Page 320: ......
Page 321: ......