Operation
Settings
Thermo Scientific
42iQHL Instruction Manual
3-95
●
Down:
●
Manual Entry:
Shows current value. If the PMT supply cell is
pressed, user can manually change the value.
●
Target Concentration:
Displays target concentration. If target
concentration cell is pressed, user can enter target concentration.
●
Calibration Concentration:
Displays calibration concentration.
●
Auto Calibration:
When pressed, the auto-calibration process is
initiated. Please allow up to 5 minutes for calibration to complete.
PMT voltage is automatically adjusted until calibration
concentration is with 1% of target concentration.
●
End Calibration:
When pressed, the auto calibration is interrupted
and the value does not change.
●
Default PMT Supply:
When pressed, the default PMT supply value
is saved.