Maintenance and Calibration
Installing the PDM Cover Panel
5-22
PDM 3600 Instruction Manual
Thermo Fisher Scientific
Figure 5–33.
PDM with cover panel removed
The front panel of the PDM may have to be removed to perform certain
maintenance tasks (
Figure 5–33
).
WARNING
Whenever the PDM case is opened for maintenance, you must
perform a case leak check after reassembling the case.
1.
Inspect the gasket and sealing surface to ensure they are clean and free
of defects. Replace the gasket, if necessary (Seal Replacement Kit: 59-
009961). Install the gasket.
2.
Install the front cover onto the PDM case. Ensure that the two screw
holes on the top of the front panel are aligned with the screw holes on
the top of the PDM.
3.
Verify that the cover gasket seats properly in place (
Figure 5–34
).
Installing the
PDM Cover Panel
To install the PDM front panel: