Basic Operation
Removing a TEOM Filter
2-6
PDM 3600 Instruction Manual
Thermo Fisher Scientific
The PDM unit is shipped with a Tapered Element Oscillating
Microbalance (TEOM) filter installed on the tapered element (TE), but the
filter must be changed after each shift.
1.
Locate the TE handle on the left side of the PDM and slide the handle
all the way to the right to unlock the mass transducer and then pull the
mass transducer out of the unit (
Figure 2–7
and
Figure 2–8
).
Figure 2–7.
Slide the PDM handle to the right
Figure 2–8.
Removing the mass transducer
Removing a
TEOM Filter
TE Handle