Maintenance and Calibration
Cleaning the Mass Transducer and Sample Lines
Thermo Fisher Scientific
PDM 3600 Instruction Manual
5-5
5.
Direct a burst of canned air into the plastic nozzle to clean the inside of
the nozzle. Use an alcohol swab to clean the inside and outside surfaces
of the plastic nozzle (
Figure 5–7
).
Figure 5–7.
Plastic nozzle
6.
Reinstall the grit pot and mass transducer.
The mass transducer and sample lines should be cleaned daily, or after each
use.
Tool/materials needed for this procedure:
●
Canned air
●
Alcohol swab
●
Small flat-head screwdriver
●
Filter exchange tool.
Cleaning the
Mass Transducer
and Sample Lines