Maintenance and Calibration
Cleaning the Mass Transducer and Sample Lines
5-6
PDM 3600 Instruction Manual
Thermo Fisher Scientific
1.
Disconnect the sample line from the cap lamp (
Figure 5–8
).
Figure 5–8.
Disconnect the sample line
2.
Remove the mass transducer from the PDM unit and remove the
TEOM filter from the mass transducer.
3.
Use canned air then an alcohol swab to clean all surfaces of the mass
transducer.
4.
Use canned air and an alcohol swab to clean the bell-shaped inlet inside
the PDM unit and then direct a stream of air through the bell-shaped
inlet (
Figure 5–9
and
Figure 5–10
).
To clean the mass transducer
and sample lines:
Disconnect Here