Basic Operation
Downloading Data
2-26
PDM 3600 Instruction Manual
Thermo Fisher Scientific
3.
When in the Download Data screen, select the sampling event that you
want to download.
4.
Select the “Download” button. If you have selected the “Auto”
downloading function, the Download Data screen may display a
“downloading progress bar” at the bottom of the screen.
5.
If the PDM unit encountered any status conditions (“errors”) during
the sample run, the “Error List” screen will display.
6.
If the error screen displays, select the “Save As” button. The “Save As”
screen will display. Accept the data file name that the software provides
for the data file or change the data file name, if necessary. Select the
“Save” button.
Note
The WinPDM software automatically names the error file with the
PDM unit's serial number (“0104”), the word “ERRORS,” and the date of
the sample run (“07-12-08”).
▲
7.
The Download Graph screen will display.
8.
When in the Download Graph screen, you can view various graphs of
the logging parameters that were recorded during the sample event.
Select the parameters to view using the drop-down menu. For more
information on the download data screen, refer to the “Download
Graph Screen” section in Chapter 4.
9.
After you have viewed the data, select the “Print Dust Data Card”
button The Dust Data Card screen will display. Select the “Print”
button to print a paper record of the sample run. This hard-copy
record will display the data entered when the sample run was first
programmed and the data results that were recorded.
10.
Select the “Save” button. The “Save As” screen will display. Accept the
data file name that the software provides for the data file or change the
data file name, if necessary. Select the “Save” button.
11.
Select the “Save” button to save the data file.