Introduction
Advanced Features
1-4
PDM 3600 Instruction Manual
Thermo Fisher Scientific
The TEOM 3600 PDM offers a number of features to provide superior
data quality, ease of use, and flexibility for the user. The following is a
partial list of the instrument’s advanced features:
●
30-minute mass concentration measurement in the range of 0.1 to 10
mg/m
3
(cumulative shift average)
●
Accuracy for mass measurement:
±
25% with 95% confidence, as
compared to gravimetric reference samplers using similar cyclones, in
the range of 0.2 mg/m
3
and greater
●
Display resolution of 0.01 mg/m
3
●
Starting with a new TEOM filter, the batteries allow a 12-hour run, up
to a loading of 5 mg
●
User-definable manual and programmed sample collection options,
based on worker shift lengths
●
Capable of indoor and outdoor use
●
Two independent internal lithium-ion battery packs (for PDM
functions and cap-lamp and auxiliary power) charge in 6 hours from a
depleted charge state (Note: the charging light may take up to 8 hours
to turn from blinking green to green as part of an enhanced charging
top-off feature)
●
Lightweight - approximately 6.2 pounds (2.812 kg)
●
Meets MSHA intrinsic safety requirements
●
Water intake through inlet does not cause hardware damage.
The WinPDM software for the TEOM 3600 PDM is a Microsoft
Windows-based program. Users should have a general understanding of
their personal computer (PC) and of the Windows operating system,
including entering and editing text, and opening, closing and saving files.