Software
85
MAXV
MAXV?
<Nx>
X X
X X Sets or returns the maximum allowed
voltage. The value should be between
0V and 15 kV otherwise the EXR
Register will be set to 5.
FRQ
FRQ?
<Nx>
X X
X X Sets or returns the target value of the
frequency. [Hz]
The range of value is within 15Hz and
400 Hz..
MODE
MODE?
INTern
EXTern
X X
X Sets or returns the state of the AC
Supply, internal or external.
STABilized
?
YES
NO
X X
X Returns "yes" if the ac source is
stabilized.
CONNectio
n:
USTA
USTB
USTAB
GSTAB
GST gA
GST gB
GST gAB
X X
X X Sets or Returns the test mode of the
system
RefCAP
INTERN?
<Nx>
X X
Returns the value of the internal
reference capacitor in [F].
EXTERN
EXTERN?
<Nx>
X X
Sets or returns the value of the external
capacitor in [F].
SWitCH
SwitCH?
INTern
EXTern
X X
X Sets or returns, if the external or the
internal capacitor is used.
TEMPeratu
re
UNIT
CELSius
FAHRenhei
t
X X
X X Sets or returns the unit of temperature
inputs. If the unit is "Celsius" all
temperature units are in "°C", if
"Fahrenheit" is set, the unit is "°F"
INSULation TEMPeratu
re
TEMPeratu
re?
<Nx>
X X
X X Sets or returns the temperature of the
insulation. The unit corresponds with the
"TEMP:UNIT" set value.