Software
71
Column Value
This data is the reference value of the limits
Column Rel
If the option
is activated “MinTol” and “MaxTol” are relative
to the reference value
, otherwise the inputs are absolute.
Column MinTol
This field indicates the minimum tolerance level of the “Pass” or
“Attention” range. Depending on the “Rel” field the value is relative or
absolute.
Column MaxTol
This field indicates the maximum tolerance level of the “Pass” or
“Attention” range. Depending on the “Rel” field the value is relative or
absolute.
Column Action
This field describes the required action when the measured value exceeds
the minimal or maximal tolerance
Button Previous
By pressing this button the previous limiter can be read.
If there is a previous cell at the SEQUENCE spreadsheet, which is marked
with a limiter sign, the limits of this cell is selected.
Button Next
By pressing this button the next limiter can be read.
If there is a next cell at the SEQUENCE spreadsheet, which is marked
with a limiter sign, the limits of this cell is selected
Tab Sheet Pass
If this tab is clicked the limiters for pass the tolerance can be edited.
You have to fill in the border of the green area.
Tab Sheet Attention
If this tab is clicked the limiters for the attention tolerance can be edited.
This is the border of the yellow area. All points located outside the
attention limits are failed (red).
Check box Relative to Reference
The minimum and maximum tolerance are relative if this box is checked,
otherwise absolute.