![Tettex MIDAS 2881 Operating Instructions Manual Download Page 73](http://html1.mh-extra.com/html/tettex/midas-2881/midas-2881_operating-instructions-manual_1092946073.webp)
Software
63
All
Signals:
Shows I
Reference
, I
Test
, U
Mains,
I
Apex
Check box Lissajous
If this option is marked, the "I Ref (Cn)" Channel is shown as a
function of the test channel "I test (Cx)", instead of as a function of
the time.
Button Difference
Activating this option will show the difference between Channel "I
Ref (Cn)" and
Channel "I test (Cx)" in a white colour.
Button Simulation
Simulated curves can be activated. Used for office version without
the Midas hardware.
Button Inc / Dec Time Base
By pressing "ms" Button, the time scale will increased. As a result,
there are more recorded data visible.
By Pressing "
s" Button, the time scale will be decreased. As a
result, there are less data will be shown.
Tab Sheet Log
This menu records all measured data in function of the time. It displays some statistical data as
Average and Standard Deviation. Be aware that the recorded data looks very instable. This is caused
by the automatic scaling of the scope, where the lowest and the highest value will be used for
minimum and maximum y-axis. At maximum 1000 values are recorded.
Summary of Contents for MIDAS 2881
Page 6: ......
Page 10: ......
Page 64: ...54 Software...
Page 109: ...Accessories and Options 99 Schematics Overview of the 5289 inductor and it s shielding...
Page 119: ...Conformity 109 13 Conformity...
Page 120: ...110 Appendix Appendix...
Page 159: ...Applications Guide 149...