![Tettex MIDAS 2881 Operating Instructions Manual Download Page 152](http://html1.mh-extra.com/html/tettex/midas-2881/midas-2881_operating-instructions-manual_1092946152.webp)
142
Applications Guide
Measurement on an ungrounded two-bushing energy storage capacitor, connection for
determination of C
G1
+ C
G2
C
1
: main
capacitor
C
G1
, C
G2
: earth insulation capacitance
Test Connections
DUT
High Voltage
to
INPUT A
to
HV GND
to
Test Mode
C
1
1
2
GND
UST A
C
G1
1
2
Tank GND
GST gA+B
C
G2
2
1
Tank GND
GST gA+B
C
G1
+ C
G2
1 + 2
Tank
GND
UST A
14.6.1
Measuring high cap values using the Resonating Inductor
With the MIDAS built-in supply, capacitances up to a maximum of C
X
= 100nF (
@
50/ 60Hz, S
max
3kVA) can be measured. For capacitances with higher values the test voltage has to be reduced or a
parallel resonating circuit can be created by using the optional resonating inductor (see also chapter
“Application Guide : Measuring high cap values using the Resonating Inductor”).
The arrangement MIDAS with resonating inductor 5288 enables measurement of capacitances up to
1
μ
F (15 kV
@
50 / 60 Hz, S
max
3kVA). For a functional description and further explanations see
chapter "Accessories and Options:Resonating Inductor".