Single Trigger
Single Trigger parameters that need to be set up include delay, samples/trigger
and VMC out.
Operating Steps:
1. Press
, then select
[
Trg Src
]
→
[
Single
]
; or press
on the front
panel directly to enable the Single Trigger. See diagram 2- 53.
Diagram 2- 53 Setting interface of Auto Trigger
2. Set the delay.
Press
[
Delay
]
to select Auto or Manual mode.
3. Set the number of samples or trigger.
Press
[
Samples/Trigger
]
to set the sample count.
4. Set the VMC Out.
The Multimeter outputs a pulse signal through the VM COMP interface on
the rear panel after the sampling signal has finished.
Press
[
VMC Out
]
to choose a Positive or Negative polarity.