98
Example 5: Limits Test
Limits Operation notifies a user if a signal has overstepped its range in
accordance to the selected High and Low Limit parameters; meanwhile the
beeper will sound an alarm (if sound is on.)
Operating Steps:
1. Press
on the front panel to select the AC Voltage measurement
function and choose a voltage range.
2. Connect the red lead to the terminal Input-HI and the black lead to the
terminal Input-LO as shown in Diagram 2-9 on page 27.
3.
Press
→
[
Limits
]
to set the Limits parameters. Press
[
Low
]
to set the
low limit value.
Diagram 3- 8 Set the Low Limit Value
4. Press
[
High
]
to set the high limit value.
Diagram 3- 9 Set the High Limit Value