![Teledyne T3DMM4-5 User Manual Download Page 53](http://html.mh-extra.com/html/teledyne/t3dmm4-5/t3dmm4-5_user-manual_1080739053.webp)
51
Dual-display Function
Dual-display function is used to improve the test and measurement functions.
Press
to open the Dual-display function and the upper right corner will
show “Dual”. By this time, press a function key if this function can be used as the
vice display, it will be displayed in the Vice Display. The Main Display will display
the function that is selected before the Dual-display function is turned on. All the
available combinations are listed in table 2-12.
Table 2- 12 Available Main/Vice Function Combinations (shade is available)
Main Display Function
DCV DCI ACV ACI FREQ PERIOD 2WR 4WR Cap
Vice
Display
Function
DCV
DCI
ACV
ACI
FREQ
PERIOD
2WR
4WR
Cap
Explanations:
1. If the same measurement function is used in both Main and Vice Display.
●
The readings in both displays will update at the same time.
●
If the math function (dBm, dB) is used in Main Display, when opening the
Vice Display, the math operation will be closed automatically. The Vice
Display will show the same measurement result as the Main Display.
●
If the math function (Statistics, Limits, Relative) is used in the Main
Display; when opening the Vice Display the result will still be shown in
Main Display. The Vice Display will show the same measurement result
as the Main Display.