![Teledyne SP Devices ADQ7DC Manual Download Page 24](http://html1.mh-extra.com/html/teledyne-sp-devices/adq7dc/adq7dc_manual_1081091024.webp)
19-2233 PC2
2019-02-01
24(50)
ADQ7DC Manual
16-1796 PC2 2019-02-01
24(50)
4.9 Level trigger
The level trigger allows data-driven acquisition. When the data on a selected channel crosses the trig-
ger level, all channels on the ADQ is triggered
1
. The level trigger is set to trigger on rising or
falling edge. Here, rising edge is illustrated.
#
DESCRIPTION
USER COMMAND
REF
a
Backplane MLVDS bus
b
Set direction for each port in the backplane
SetDirectionMLVDS
c
Output: Select output port for trigger output
SetupTriggerOutput
d
Output: This is the source for the trigger output signal
SetupTriggerOutput
e
Input: Select port for trigger sources
SetTriggerMaskMLVDS
f
Input: The backplane trigger signal is OR:ed from the selected ports
g
Input: this is the backplane trigger to the trigger module
Figure 15: MLVDS port 17 - 20 as trigger signals.
1. Firmware option FWPD offers a channel independent level trigger. See specific documentation in form of user
guides
and application note
.
' (
'(
'(
'(
'(
)
% *
+,
-./
'(
'(
+,
./
+,
-.0
+,