![Teledyne SP Devices ADQ7DC Manual Download Page 29](http://html1.mh-extra.com/html/teledyne-sp-devices/adq7dc/adq7dc_manual_1081091029.webp)
19-2233 PC2
2019-02-01
29(50)
ADQ7DC Manual
16-1796 PC2 2019-02-01
29(50)
4.11.5 Distributing level trigger
Multiple boards may be triggered by a level trigger on one channel in one of the digitizers. This mode is
intended for systems with a reference event on one channel that starts the acquisition on all channels.
It may also be used for designing a high precision trigger. The trigger event on channel D in
is recorded on the channel D. The trigger timing is then calculated with high precision using interpola-
tion.
#
DESCRIPTION
USER COMMAND
a
Select a channel as level trigger.
SetupLevelTrigger
b
Select level trigger as trigger out.
SetupTriggerOutput
c
Internal hardware connection trigger out to trigger in (shared SMA
connector).
d
Select external trigger for triggering
SetTriggerMode
e
External cable connection to other digitizers.
f
Select external trigger for triggering.
SetTriggerMode
Figure 21: Distributing level trigger.
'(
!)*"
!)*"
!
$
+
,
+-
.
!
/
! *
/ * *
'(
!)*"
!)*"
!
$
+
,
+-
.
!
'(
'0(
'!(
'(
/
!
-*
,