Monitoring the SDI Physical Layer
2.
To reset both the SMPTE 259 (SD)
and SMPTE 292 (HD) monitoring alarm
thresholds back to the factory-default
values, touch the
Set to SMPTE259/292
Limits
soft key.
Setting the SMPTE limits also sets both
clock recovery circuits to measure both
timing and alignment jitter. This can be
seen on the SDI Status display.
3.
To set individual threshold limits, touch
the
Show Limits
soft key (in the Settings
submenu) to enable displaying of the
limit values. Which threshold limits
are displayed depends on the installed
options.
4.
Touch the limit value to set and use the
general-purpose knob to change the
value.
NOTE.
You can use the SELECT front-panel
button to sequentially step through and
select the threshold limit values.
5.
Touch the
Close Settings
soft key.
Setting the Equalizer Bypass
1.
Touch the
Settings
soft key. (See
page 86,
Displaying the Eye
Measurement Menu
.)
Quick Start User Manual
89