Specifications
TDS6000 Series Service Manual
1- 19
Table 1- 12: Certifications and compliances
Category
Standards or description
EC Declaration of Conformity --
EMC
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was
demonstrated to the following specifications as listed in the Official Journal of the European Union:
EN 61326 Emissions
1, 3
Class A Radiated and Conducted Emissions
EN 61326 Immunity
1
,
2
,
4
IEC 1000-4-2
Electrostatic Discharge Immunity
±
4 kV contact discharge,
±
8 kV air discharge,
performance criterion B
IEC 1000-4-3
RF field immunity
3 V/m, 80 MHz to 1 GHz,
80% amplitude modulated with a 1 kHz sinewave
performance criterion A
IEC 1000-4-4
Electrical Fast Transient/Burst Immunity
1 kV on AC mains, 500 V on !/O cables, performance criterion B
IEC 1000-4-5
AC Surge Immunity
1 kV differential mode, 2 kV common mode,
performance criterion B
IEC 1000-4-6
RF Conducted Immunity
3 V, 150 kHz to 80 MHz, amplitude modulated with a
1 kHz sinewave, performance criterion A
IEC 1000-4-11
AC Mains Voltage Dips and Interruption Immunity
100% reduction for one cycle, performance criterion B
EN 61000-3-2
Power Harmonic Current Emissions
1
If interconnect cables are used, they must be low-EMI shielded cables such as the
following Tektronix part numbers or their equivalents: 012-0991-01, 012-0991-02 or
012-0991-03 GPIB Cable; 012-1213-00 (or CA part number 0294-9) RS-232 Cable;
012-1214-00 Centronics Cable; or LCOM part number CTL3VGAMM-5 VGA Cable.
2
The performance criteria for when the oscilloscope is subjected to the conditions
described above are defined as follows:
A — 10 mV/division to 1 V/division:
≤
0.4 division waveform displacement or
≤
0.8
division
increase in peak-to-peak noise
5 mV/division and 2 mV/division, typical:
≤
8 mV increase in peak-to-peak noise
B — temporary, self-recoverable degradation or loss of performance is allowed, but no
change of actual operating state or loss of stored data is allowed
C — temporary loss of function is allowed provided that the function is self recoverable
or can be restored by the operation of the controls
3
Radiated emissions may exceed the levels specified in EN 61326 when this
oscillo-
scope
is connected to a test object.
4
USB mouse and keyboard only, performance criterion C. Normal USB keyboard or
mouse operation can be restored by unplugging and reconnecting the USB connector
to the
oscilloscope
.
Australia/New Zealand Declara-
tion of Conformity — EMC
Complies with EMC provision of Radiocommunications Act per the following standard(s):
AS/NZS 2064.1/2 Industrial, Scientific, and Medical Equipment: 1992
Summary of Contents for TDS6404
Page 4: ......
Page 14: ...Table of Contents x TDS6000 Series Service Manual...
Page 18: ...Service Safety Summary xiv TDS6000 Series Service Manual...
Page 52: ...Operating Information 2 12 TDS6000 Series Service Manual...
Page 56: ...Theory of Operation 3 4 TDS6000 Series Service Manual...
Page 60: ...Performance Verification 4 4 TDS6000 Series Service Manual...
Page 84: ...Performance Tests 4 28 TDS6000 Series Service Manual...
Page 150: ...Adjustment Procedures 5 2 TDS6000 Series Service Manual...
Page 156: ...Maintenance 6 6 TDS6000 Series Service Manual...
Page 170: ...Removal and Installation Procedures 6 20 TDS6000 Series Service Manual...
Page 228: ...Repackaging Instructions 6 78 TDS6000 Series Service Manual...
Page 232: ...Options 7 4 TDS6000 Series Service Manual...
Page 234: ...Electrical Parts List 8 2 TDS6000 Series Service Manual...
Page 252: ...Mechanical Parts List 10 16 TDS6000 Series Service Manual...