Performance Tests
4- 26
TDS6000 Series Service Manual
TDS6000 Test Record (cont.)
Instrument Serial Number:
Certificate Number:
Temperature:
RH %:
Date of Calibration:
Technician:
TDS6000 performance test
Maximum
Outgoing
Incoming
Minimum
Time base system
Long term sample rate, delay time, and internal
reference accuracy
9999.975 kHz
__________
__________
10000.025 kHz
Delta time measurement
Ch 1
N/A
__________
__________
≤
0.015 ns
Ch 2
N/A
__________
__________
≤
0.015 ns
Ch 3
N/A
__________
__________
≤
0.015 ns
Ch 4
N/A
__________
__________
≤
0.015 ns
Trigger system accuracy
Time accuracy for pulse, glitch, timeout, and
Width, Hor. scale
≤
1
s
Lower Limit
Upper Limit
3.5 ns
3.5 ns
__________
__________
__________
__________
6.5 ns
6.5 ns
Time accuracy for pulse, glitch, timeout, and
width, Hor. scale > 1
s
Lower Limit
Upper Limit
1.9
s
1.9
s
__________
__________
__________
__________
2.1
s
2.1
s
Probe compensation output signal
Frequency
950 Hz
__________
__________
1.050 kHz
Voltage (difference)
160 mV
__________
__________
240 mV
Summary of Contents for TDS6404
Page 4: ......
Page 14: ...Table of Contents x TDS6000 Series Service Manual...
Page 18: ...Service Safety Summary xiv TDS6000 Series Service Manual...
Page 52: ...Operating Information 2 12 TDS6000 Series Service Manual...
Page 56: ...Theory of Operation 3 4 TDS6000 Series Service Manual...
Page 60: ...Performance Verification 4 4 TDS6000 Series Service Manual...
Page 84: ...Performance Tests 4 28 TDS6000 Series Service Manual...
Page 150: ...Adjustment Procedures 5 2 TDS6000 Series Service Manual...
Page 156: ...Maintenance 6 6 TDS6000 Series Service Manual...
Page 170: ...Removal and Installation Procedures 6 20 TDS6000 Series Service Manual...
Page 228: ...Repackaging Instructions 6 78 TDS6000 Series Service Manual...
Page 232: ...Options 7 4 TDS6000 Series Service Manual...
Page 234: ...Electrical Parts List 8 2 TDS6000 Series Service Manual...
Page 252: ...Mechanical Parts List 10 16 TDS6000 Series Service Manual...